Категории

MSP Series

Features

  • Universal AC input / Full range
  • Medical safety approved (2xMOOP)
  • Withstand 300VAC surge input for 5 seconds
  • Built-in active PFC function
  • Protections:  
    • Short circuit / Overload / Over voltage /  
    • Over temperature
  • 1U low profile (MSP-100/200/300/450)
  • Built-in constant current limiting circuit
  • Built-in remote sense function (MSP-200/300/450/600)
  • No load power consumption <0.5W (<0.6W for MSP-450 ; <0.8W for MSP-600) 
  • Built-in current sharing (MSP-600-24/36/48)
  • Built-in remote ON/OFF control 
  • Built-in 5V/0.3A standby output (MSP-200/300/450/600)
  • Built-in DC OK signal (MSP-300/450/600)
  • LED indicator for power on
  • 5 years warranty

Model No .

MSP - 100

MSP - 200

MSP - 300

MSP - 450

MSP - 600

AC input voltage range

85~264VAC; 120~370VDC

AC inrush current (max . )

Cold start, 65A at 230VAC

Cold start, 70A at 230VAC

Cold start, 80A at 230VAC

DC adjustment range

±15% rated output voltage

Overload protection

Range

105%~135%

Type

Constant current limiting, auto - recovery

Over voltage protection

115%~145%

Withstand voltage

I/P - O/P: 4kVAC, I/P - FG: 2kVAC, O/P - FG: 0 . 5kVAC, 1 minute

Working temperature

-40~+60°C

-40~+70°C (refer to output derating curve)

Safety standards

ANSI/AAMI ES60601 - 1, IEC60601 - 1 approved

EMC standards

EN55011 class B, EN61000 - 4 - 2,3,4,5,6,8,11, EN61000 - 3 - 2,3, EN60601 - 1 - 2

Connection

7P / 9 . 5mm pitch terminal block with cover

7P / 11mm pitch terminal block with cover

3+6P / 10 &11mm pitch terminal block with cover

Dimension (LxWxH)(mm)

159x 97x 38

199x 98x 38

199x 105x 41

218x 105x 41

218x 105x 63 . 5

Case No .

901 I

902E

980A

995A

977A